ANALIZADOR DE PUNTO DE FLAMA DE COPA CERRADA DE MICROESCALA CONTINUA AUTOMÁTICA

Trademark:
Koehler Instrument Company, Inc.

Temperature Range:
-30 °C to 405°C (-22 °F to 788 °F) below 0 °C an external cryogenic recirculatory is required

Heating Rate:
2.5 °C/min ± 0.3 °C/min or 5.5 °C/min ± 0.5 °C/min; can be customized in the range of 0.5 to 12 C/min

Temperature Accuracy:
± 0.1 °C

Pressure Range:
0 kPa to 200 kPa

Stirring Rate:
Can be customized from 50 RPM to 300 RPM

Categoría: Etiqueta:

Descripción

Under specified test conditions, when a sample is heated to the temperature called flash point, the sample vapor and air mixture are ignited instantly. Flash Point is one of the indicators for the evaluation of a materials’ combustion tendency. If the flash point Is low, the combustion risk would be high. As for transport and safety regulation, flash point is used for the classification of “flammable liquids”. Also, flash point is used for quality control index or determining the existence of highly volatile or flammable substances in relatively nonvolatile or nonflammable matter, e.g. lubricant mixed with a small amount of diesel or gasoline.

The Koehler K24880 Microscale Continuously Closed Cup Flash Point Tester works in the following manner. The temperature of the analyzer is adjusted to at least 18 °C below the expected flash point. The sample is injected into the sample cup. The operator must ensure that the sample and sample cup temperature are both at least 18 °C below the expected flash point (the sample and sample cup may be cooled if necessary). Then the sample cup is raised towards the oven forming a closed test chamber, which is not sealed.

After every ignition, the proper volume of air goes into the test chamber to supply the next ignition with oxygen to burn. The pressure in the unsealed closed test chamber should be ambient atmospheric pressure, except during the short period of air introduction and the time when the flash point is achieved. The pressure increase comes after discharge of air, so each time after the air discharges, the chambers pressure is higher than ambient atmospheric pressure. When the pressure increase exceeds the specified limit, this is recorded as the flash point temperature (which has not been corrected).