SISTEMAS DE REFRIGERACIÓN DE FILTRACIÓN DUAL DE PRUEBA DE REMOJO EN FRÍO

Trademark:
Koehler Instrument Company, Inc.

Conforms to the specifications of:
ASTM D1657; IP 235; ISO 3993

Controller Sensitivity:
±0.5°F (±0.2°C)

Capacity:
two (2) K26150 cylinders

Categoría: Etiqueta:

Descripción

ressure Hydrometer Cylinder Transparent plastic cylinder mounted between machined aluminum end plates and surrounded by stainless steel safety guard. Use together with ASTM 310H Thermohydrometer to determine density or relative density of LPG and light hydrocarbons. Equipped with inlet, outlet and vapor vent valves for admitting sample and purging cylinder. End plates have positive sealing buna-N O-rings and are joined by sturdy steel support rods. Top plate detaches easily without tools for insertion or removal of thermohydrometer. Safety relief valve prevents unsafe pressure build-up inside cylinder. Mounted on a finished steel base. Constant Temperature Water Bath Immerses two Pressure Hydrometer Cylinders at 60°F (15°C) for density and relative density determinations of LPG and other light hydrocarbons. Mechanically refrigerated cooling system maintains sub-ambient temperature. Thermistor activated solid state temperature controller and 750W copper immersion heater maintain bath temperature with ±0.5°F (±0.2°C) stability. A 1⁄20 hp ball bearing stirrer circulates the bath medium to assure temperature uniformity. Stainless steel tank is fiberglass insulated. Equipped with overflow standpipe/drain. Steel exterior has a durable polyurethane enamel finish.